G2/132 Gas Permeability Tester
Feature
Professional
- The system provides proportion and standard test modes with convenient parameter settings
- The gas transmission rate as well as the coefficients of permeability, solubility and diffusion can be obtained at one operation
- 3 identical specimens can be tested simultaneously with the average value as test result
- Wide test range for different materials with high, medium and low barrier properties
- Various types of gases are testable: sole gas, mixed gases, poisonous gases, explosive gases and other dangerous gases (customization is required)
- World exclusive data fitting function that could easily calculate gas permeability and other parameters at different temperatures
- Top quality parts and components made by world famous brands are used to ensure reliable overall product performance
- Reference film calibration ensures accurate and universal test data
High-end
G2/132 utilizes Labthink's latest embedded computer control system that provides a better performance than traditional single chip system.
- Patented integrated design of three test cells improves the test efficiency and reduces the space occupancy of the instrument
- Embedded computer control system provides safer and more reliable data management as well as test operation
- The instrument can be easily operated with a mouse, a keyboard, and a monitor, without requiring a PC
- The system is equipped with four USB ports and dual Internet ports for convenient data transmission
Intelligent
The instrument is equipped with Labthink's latest intelligent operating software, with user-friendly operating interface and intelligent data management. It also supports LystemTM Lab Data Sharing System, which ensures uniform management of test results and test reports.
- Status monitoring and intelligent reminding of sensor calibration ensure instrument in the best working condition
- The system automatically calculates the statistical information of instrument utilization rate and test times
- Embedded help document for user viewing at any time
- Multi-level account control for better data management and protection
- The system utilizes embedded data saving technology to save detailed information and provide convenient and various searching and viewing functions
- One time value input and the system automatically gives data comparison after each test
- Supports LystemTM Lab Data Sharing System for uniform and systematic data management
Spesification
Technical Specifications
Specifications | Film Test |
---|---|
Test Range | 0.05~20,000 cm3/m2·24h·0.1MPa |
Temperature Range | 15°C~55°C (room temperature 23°C) |
Temperature Accuracy | ±0.1°C (standard) |
Humidity Range | 0%RH, 2%~98.5%RH, 100%RH (humidity generator is optional) |
Humidity Accuracy | ±1%RH |
Vacuum Resolution | 0.1 Pa |
Vacuum Degree of Test Chamber | |
Number of Specimens | 3 |
Specimen Size | Φ97 mm |
Test Area | 115.44 cm2 |
Test Gas | O2, N2 and CO2 (outside of supply scope) |
Test Pressure | -0.1MPa~+0.1MPa |
Gas Supply Pressure | 0.4 MPa~0.6 MPa |
Port Size | Φ6 mm PU Tubing |
Instrument Dimension | 690 mm (L) x 350 mm (W) x 360 mm (H) |
Power Supply | AC (85~264)V (47~63)Hz |
Net Weight | 54 kg |
Inquiry
Video
Tidak ada video untuk produk ini