Temp/RH 2.5% Data Logger. HOBO - UX100-011
Temp/RH 2.5% Data Logger
The HOBO UX100-011 Temperature/Relative Humidity data logger records temperature andrelative humidity (within 2.5% accuracy) in indoor environments with its integrated sensors. This data logger is ideal for monitoring occupant comfort in office buildings, tracking food storage conditions in warehouses, logging temperature trends in server rooms, and measuring humidity levels in museums.
Environment:
The UX100-011 is for use in Indoorenvironments
Measurements:
The UX100-011 supports the following measurements:
Temperature • Relative Humidity
IMPORTANT:
HOBO data loggers requires HOBOware software (USB interface cable included w/CD purchase) and a communications device. See compatible items below.
The HOBO UX100-011 Temperature/Relative Humidity data logger records temperature andrelative humidity (within 2.5% accuracy) in indoor environments with its integrated sensors. This data logger is ideal for monitoring occupant comfort in office buildings, tracking food storage conditions in warehouses, logging temperature trends in server rooms, and measuring humidity levels in museums.
Environment:
The UX100-011 is for use in Indoorenvironments
The UX100-011 supports the following measurements:
Feature
- Easy-to view LCD display
- Large memory capacity
- Visual high & low alarm thresholds
- New Burst and Statistics logging modes
- User-replaceable Relative Humidity sensor
Spesification
To see the full specifications for this product, please see our product manual found under the documentation tab.
Temperature Sensor
Range: -20° to 70°C (-4° to 158°F)
Accuracy: ±0.21°C from 0° to 50°C (±0.38°F from 32° to 122°F), see Plot A
Resolution: 0.024°C at 25°C (0.04°F at 77°F); see Plot A
Response Time: 4 minutes in air moving 1 m/s (2.2 mph)
Drift: <0.1°C (0.18°F) per year
RH Sensor
Range: 1% to 95%
Accuracy: ±2.5% from 10% to 90% typical to a maximum of ±3.5% including hysteresis, see Plot B
Resolution: 0.05% at 25°C (77°F)
Response Time: 11 seconds to 90% in airflow of 1 m/s (2.2 mph)
Drift: <1% per year typical
Logger
Logger Operating Range:
Logging: -20° to 70°C (-4° to 158°F); 0 to 95% RH (non-condensing)
Launch/Readout: 0° to 50°C (32° to 122°F) per USB specification
Logging Rate: 1 second to 18 hours, 12 minutes, 15 seconds
Logging Modes: Normal, Burst or Statistics
Memory Modes: Wrap when Full or Stop when Full
Start Modes: Immediately, Push Button, Date & Time or Next Interval
Stop Modes: When Memory Full, Push Button, or Date & Time
Restart Mode: Push Button
Time Accuracy: ±1 minute per month at 25°C (77°F), see Plot C
Battery Life: 1 year, typical with logging rate of 1 minute and sampling interval of 15 seconds or greater
Battery Type: One 3V CR2032 lithium battery
Memory: 128 KB (84,650 measurements, maximum)
Download Type: USB 2.0 interface
Full Memory Download Time: 20 seconds
LCD: LCD is visible from: 0° to 50°C (32° to 122°F); the LCD may react slowly or go blank in temperatures outside this range
Size: 3.66 x 8.48 x 2.29 cm (1.44 x 3.34 x 0.9 in.)
Weight: 30 g (1.06 oz)
Environmental Rating: IP50
Inquiry
Video
Tidak ada video untuk produk ini